COMPANY HEADER
Request Information
Versatile total hydrocarbon analysis down to ppb levels.
Unit of Measure

Specifications

Brands

AMETEK® MOCON®

Industry Focus

Specialty Gas Oil and Gas Environmental Semiconductors

Technical Specifications

Detector

Flame Ionization Detector (FID)

Ranges (CH4)

User definable based upon calibration:
  • Very Low - 0.01 ppm to 200 ppm
  • Low - 0.1 ppm to 2,000 ppm
  • Medium - 0.3 ppm to 20,000 ppm
  • High - 0.003% ppm to 100% ppm

Accuracy, Repeatability

+1% full-scale response

Drift, Zero

+0.025% full-scale over 24 hours

Drift, Span

+1% full-scale, 24 hours

Response Time

T90 < 5 seconds

Sampling

Internal single or multi-point, (pump optional)

Alarms

Multilevel concentration and fault alarms, audible and visual

Calibration

Programmable automatic or manual calibration
Support Gas Requirement1
  • UHP Hydrogen (H2) 30 cc/min
  • Zero air - 175 cc/min
  • Span gas - methane is typical

Display Type

Graphical Touch Display

Display Size

7 in17.8 cm

Display Orientation

Diagonal

Output

  • LAN/Ethernet, RS-485
  • Optional outputs
    • Analog, programmable 0-20 mA/4-20 mA isolated output; Relays, 5 programmable From A relays rated to 3 A at 230 VAC
    • If (a) is installed, 8 additional analog outputs and up to 16 additional relays can be added

Operating Temperature

32 to 104 ºF0 to 40 ºC

Non-Condensing Operating Humidity

0 to 95 %

Configuration

Benchtop or 19" (48.3 cm) Rack-Mount, 3U

Physical Specifications

Height

5.25 in13.02 cm

Width

19 in48.26 cm

Depth

16.25 in41.28 cm

Weight

< 20 lb9.07 kg

Electrical Specifications

Alternating Current (AC) Voltage

100 to 240 V

Frequency

50 Hz60 Hz

Current

1 A

Accessories

Accessories

Preventative Maintenance Kit
  • Includes all necessary parts
Gas Generators
  • Hydrogen generator
  • Zero air generators Service & Support Options
  • Start-up and training, one day onsite

Unique Needs

Unique Needs

Specialty gas producers require robust quality control and quality assurance to keep products consistent and customers happy. Accurate trace level measurements and a reliable instrument are vital to maintaining tight operations.

For oil and gas companies, downtime means major financial loss. That's why a rugged instrument with user-focused diagnostics and troubleshooting is essential.

Whether collecting ambient air quality data or monitoring emissions to stay below required levels, environmental applications require highly accurate results and easy access to data for compliance and reporting.

Features

Features

High Performance Effortless Operation

Industry-approved gas measurement

Reliable performance meets modern interface
The BASELINE® Series of Continuous Hydrocarbon Analyzers (CHAs) from AMETEK MOCON® are instruments designed for continuous ambient or process hydrocarbon gas measurement in environmental or industrial settings. All instruments in the series are powered by our high-quality Flame Ionization Detection (FID) sensor, which enables fast, accurate measurement and low detection limits with minimal interference.

The CHA Series combines reliable, industry-proven technology with a modern interface and straightforward troubleshooting, allowing you to focus on the results.

Operation designed for humans
High-quality data isn't enough if the instrument is difficult to use. The CHA Series features an intuitive, streamlined interface. Instruments include a range of capabilities to help simplify your process, including easily read results, real-time graphing, the ability to review historical data without pausing instrument operation, and dual display output. The CHA Series also features networking capabilities for easier connectivity, data sharing within a network, and remote diagnostics.

Fast and Accurate - FID
Our proprietary FlowGuard electronic control regulates the delivery of fuel, air and a small portion of the sample gas to the FID. During combustion, gases are ionized, measured and reported as a concentration. This method provides fast gas measurement with some of the highest accuracy levels on the market.

Intuitive Interface
  • Modern GUI interface
  • Dual display capability
Simple Diagnostics
  • On-board data storage
  • Real-time graphing
  • Synchronous operation and data review
High Performance
  • Low detection levels
  • Fast measurement
  • Automatic calibration
  • 1 Fuel blend options available