Items |
/Asset/BASELINE__Series_Continuous_Hydrocarbon_Analyzers.png THA BASELINE® Series Versatile Total Hydrocarbon (THA) Analyzer |
/Asset/BASELINE__Series_Continuous_Hydrocarbon_Analyzers.png MNME BASELINE® Series Methane/Non-Methane (MNME) Hydrocarbon Analyzer |
/Asset/BASELINE__Series_Continuous_Hydrocarbon_Analyzers.png TCA BASELINE® Series Hydrocarbons and CO2 Total Carbon (TCA) Analyzer |
|||
Description | N/A Versatile total hydrocarbon analysis down to ppb levels. | N/A Oxidation catalyst allows precise measurement of methane and non-methane gases. | N/A Catalytic methanizer allows accurate reading of combined CO/CO2 measurement | |||
Brands | N/A AMETEK® MOCON® | |||||
Detector | N/A Flame Ionization Detector (FID) | N/A Flame Ionization Detector (FID) with Oxidizer | N/A Flame Ionization Detector (FID) with Methanizer | |||
Ranges (CH4) |
N/A
User definable based upon calibration:
|
N/A
User definable based upon calibration:
|
N/A
User definable based upon calibration:
|
|||
Accuracy, Repeatability | N/A +1% full-scale response | N/A +1% full-scale | N/A +1% full-scale | |||
Drift, Zero | N/A +0.025% full-scale over 24 hours | N/A +0.01% full-scale, 24 hours | N/A +0.01% full-scale, 24 hours | |||
Drift, Span | N/A +1% full-scale, 24 hours | |||||
Response Time | N/A T90 < 5 seconds | N/A T90 < 30 seconds | N/A T90 < 30 seconds | |||
Sampling | N/A Internal single or multi-point, (pump optional) | N/A Internal single or multi-point, (pump optional) | N/A Internal single point (pump optional) | |||
Alarms | N/A Multilevel concentration and fault alarms, audible and visual | |||||
Calibration | N/A Programmable automatic or manual calibration | |||||
Support Gas Requirement1 |
N/A
|
N/A
|
N/A
|
|||
Display Type | N/A Graphical Touch Display | |||||
Display Size | N/A 7 in17.8 cm | |||||
Display Orientation | N/A Diagonal | |||||
Output |
N/A
|
|||||
Operating Temperature | N/A 32 to 104 ºF0 to 40 ºC | |||||
Non-Condensing Operating Humidity | N/A 0 to 95 % | |||||
Configuration | N/A Benchtop or 19" (48.3 cm) Rack-Mount, 3U | |||||
Height | N/A 5.25 in13.02 cm | |||||
Width | N/A 19 in48.26 cm | |||||
Depth | N/A 16.25 in41.28 cm | |||||
Weight | N/A < 20 lb9.07 kg | |||||
Alternating Current (AC) Voltage | N/A 100 to 240 V | |||||
Frequency | N/A 50 Hz60 Hz | |||||
Current | N/A 1 A | |||||
Industry Focus | N/A Specialty Gas Oil and Gas Environmental Semiconductors | N/A Environmental Semiconductors | N/A Specialty Gas | |||
Accessories |
N/A
Preventative Maintenance Kit
|
|||||
Unique Needs |
N/A
Specialty gas producers require robust quality control and quality assurance to keep products consistent and customers happy. Accurate trace level measurements and a reliable instrument are vital to maintaining tight operations. For oil and gas companies, downtime means major financial loss. That's why a rugged instrument with user-focused diagnostics and troubleshooting is essential. Whether collecting ambient air quality data or monitoring emissions to stay below required levels, environmental applications require highly accurate results and easy access to data for compliance and reporting. |
N/A Whether collecting ambient air quality data or monitoring emissions to stay below required levels, environmental applications require highly accurate results and easy access to data for compliance and reporting. | N/A Specialty gas producers require robust quality control and quality assurance to keep products consistent and customers happy. Accurate trace level measurements and a reliable instrument are vital to maintaining tight operations. | |||
Features |
N/A
High Performance Effortless Operation Industry-approved gas measurement Reliable performance meets modern interface The BASELINE® Series of Continuous Hydrocarbon Analyzers (CHAs) from AMETEK MOCON® are instruments designed for continuous ambient or process hydrocarbon gas measurement in environmental or industrial settings. All instruments in the series are powered by our high-quality Flame Ionization Detection (FID) sensor, which enables fast, accurate measurement and low detection limits with minimal interference. The CHA Series combines reliable, industry-proven technology with a modern interface and straightforward troubleshooting, allowing you to focus on the results. Operation designed for humans High-quality data isn't enough if the instrument is difficult to use. The CHA Series features an intuitive, streamlined interface. Instruments include a range of capabilities to help simplify your process, including easily read results, real-time graphing, the ability to review historical data without pausing instrument operation, and dual display output. The CHA Series also features networking capabilities for easier connectivity, data sharing within a network, and remote diagnostics. Fast and Accurate - FID Our proprietary FlowGuard electronic control regulates the delivery of fuel, air and a small portion of the sample gas to the FID. During combustion, gases are ionized, measured and reported as a concentration. This method provides fast gas measurement with some of the highest accuracy levels on the market. Intuitive Interface
|
|||||
|