Brands |
N/A AMETEK® Dansensor® MOCON® |
Available Configurations |
N/A Dansensor® MAP Check 3 Vacuum Analyzer |
Oxygen Measurement (Zirconia Sensor) |
N/A Standard in all models |
Carbon Dioxide Measurement (Dual Beam Infrared) |
N/A Model dependent |
Control of Modified Atmosphere Packages (MAP) Mix Provectus |
N/A Yes |
Number of Test Programs (Devices with Liquid Crystal Display (LCD) Display) |
N/A Up to 1,000 |
Measuring Range1 | N/A 0 to 100 % |
Resolution for O2 |
N/A <1%, 0.001 <10%, 0.01 O2: ==10%, 0.1 |
Resolution for CO2 |
N/A CO2: 0-100%, 0.1 |
Sensor Accuracy for O2 |
N/A O2: =0.1%, =0.1%, ±1% of reading; <0.1%, ±0.001% absolute |
Sensor Accuracy for CO2 |
N/A CO2: 0-100%, ±1.5% of reading + 0.5% absolute |
Note for Sensor Accuracy |
N/A System accuracy depends on die geometry and machine timing |
Heating Time |
N/A 10 min |
Models |
N/A Available with Liquid Crystal Display (LCD) display or as "Black Box" without display |
Connections |
N/A 2 x RS232C| Local Area Network (LAN) 10/100 Mbit (Modbus TCP)| Universal Serial Bus (USB)| Host for Import/Export| 24 VDC Logic for Machine Control Signals and Alarms |
Height |
N/A 192 mm |
Width |
N/A 230 mm |
Depth |
N/A 375 mm |
Weight2 | N/A 8.5 to 11.5 kg |
Alternating Current (AC) Voltage3 | N/A 103 to 132 V207 to 264 V |
Frequency |
N/A 47 to 63 Hz |
Compliances |
N/A European Conformity (CE) |
Protection Kit |
N/A Ingress Protection (IP45) National Electrical Manufacturers Association (NEMA 3S) |
Bracket, Assembly |
N/A Can be combined with MAP Mix Provectus and MAP Check 3: 2 brackets, 8 screws. |
Buffer Tank Measurement |
N/A Model Dependent |
Benefits |
N/A
|
Features |
N/A
|
How Does it Work? |
N/A
|
Additional Information |
N/A
A Win-Win For Thermoforming And Tray Sealing Lines Automatic testing with the Dansensor® MAP Check 3 Vacuum is not just faster and more reliable than manual testing, it is also more efficient. If there is a problem with the gas combination, the system notifies the operator immediately, saving time. If preset limits are exceeded, the analyzer simply stops the process, which means no wasted products or packaging. |