Brands |
N/A AMETEK® MOCON® OX-TRAN® PERMATRAN-W® |
Conformance Standards |
N/A European Conformity (CE) Safety Compliance |
Connectivity |
N/A Dual RS 485 Ports |
Depth |
N/A 26 in66.4 mm |
Width |
N/A 12 in30.4 mm |
Height |
N/A 12 in30.4 mm |
Weight |
N/A 63 lb28.6 kg |
Temperature Control Range |
N/A 20 to 50 ºC |
Temperature Control Range Tolerance (±) |
N/A 0.5 ºC |
Humidity Control, Film Cell, Test Gas |
N/A 100 % |
Film Resolution |
N/A 1 g/(m²·day) |
Film Repeatability |
N/A 1 % |
Test Range for Film Unmasked (10 cm²) |
N/A 500 to 100000 g/(m²·day)32.5 to 6500 g/(100 in²·day) |
Industry Standards |
N/A American Society for Testing and Materials (ASTM) D 6701 |
Features |
N/A
MOCON®’s WinPerm™ Permeability software is designed for speed and ease of use. Features include:
|
Accuracy at High Transmission Levels |
N/A Obtaining accurate and repeatable results when testing highly permeable materials is a challenge. Traditional Gravimetric methods such ASTM E96 have issues with loss of driving force due to the air gap between the moisture source and the sample resulting in erroneous and inconsistent test results. The 101K WVTR instrument utilizes a modified inverted water cup concept to eliminate the air gap problem and maintain constant 100% RH driving force throughout the test period. The precision RH sensor on the carrier gas side provides accurate results. The automated system maintains consistent conditions throughout the tests yielding repeatable results without operator intervention. We also offer a special calibration film for the instrument to ensure accurate test results. |
Applications |
N/A The 101K is ideal for measuring WVTR in materials such as porous and breathable polymer membranes, thin paper with polymer coatings and thin polymer nonwovens (with non-textured surface). Such barriers are common applications in the healthcare industry as diapers and bandages and in the construction industry as moisture barriers (house wraps) and soil control. |
High Throughput |
N/A
|
Simple User Interface |
N/A
|
Consistent Conditions |
N/A The unique inverted cup design allows transmission rate testing without an air gap that maintains a constant 100% RH driving force of water vapor throughout the test without operator interaction. The air gap that exists in traditional test methods negatively effects the RH at the barrier sample that reduces the accuracy of the test. When testing high transmitting barriers, the large amount of moisture transmitting through the barrier can result in changing the RH driving force as the test progresses. This inverted cup design events changes in driving force thus ensuring accurate and repeatable test results. |
Porosity vs Permeation |
N/A Additionally, the inverted cup design relies on testing with equal pressure on both sides of the sample allowing accurate measurement of the permeation transmission rate without the need for pressure differential. These results are true permeation via molecular diffusion. This equal pressure condition eliminates any artifact of water vapor transfer through the porosity of the barrier and provides a true measure of the breathability of the barrier. |
Simple Operation |
N/A
Place all desired sample films in position, attach the test head and adjust the flow to the desired level. Enter the conditions such as temperature into the connected computer with AMETEK MOCON®’s WinPerm™ software and start the test. Monitor each cell from the computer to see progress and test results when complete. Digital pressure and flow controls allow for simple “set and forget” test conditions. A single water reservoir in the head achieves 100% RH conditions throughout the test for all 6 cells. ASTM D6701 cites the 100K, predecessor of the 101K, as the only known instrument to comply with the method. The 101K provides the most accurate and repeatable permeation results for your high transmitting barriers. |
Innovative Relative Humidity (RH) Control |
N/A
|
Accurate Results |
N/A
|
Additional Information |
N/A
Time Savings By measuring moisture concentration, carrier gas flow and temperature, the 101K is easy to set up and requires no supervision through the test. The fully automated test provides rapid test results to help reduce overall product development time or shorten product quarantine times in the production environment. Additionally, the instrument can test up to six samples simultaneously, ideal for a QC lab supporting production operations. The PERMATRAN-W® Model 101K provides true and accurate water vapor permeation rates for breathable membranes or films. |