The 9000 Heated analyzer functions likeThe 9000 THA but is configured for single point analysis (with or without a sample pump) of samples heated up to 376 °F (191 °C) for pre-filtered (< 10 microns) noncondensing samples.
Unit of Measure

Specifications

Brands

N/A AMETEK® MOCON®

Detector

N/A Flame Ionization Detector (FID) with Oxidizer

Ranges (CH4)

N/A User definable based upon calibration:
  • Very Low - 0.15 ppm to 200 ppm
  • Low - 0.3 ppm to 2,000 ppm
  • Medium - 0.6 ppm to 20,000 ppm
  • High 0.003% to 50%

Accuracy

N/A +1% full-scale response.

Drift, Zero

N/A +0.025% full-scale, 24 hours.

Drift, Span

N/A +1% full-scale, 24 hours.

Response Time

N/A T90 < 5 s

Sampling

N/A Internal single point (pump optional)

Alarms

N/A Multilevel concentration and fault alarms, audible and visual.

Calibration

N/A Manual Calibration Programmable Automatic
Support Gas Requirement1 N/A
  • UHP Hydrogen (H2) 30 cc/min
  • Zero air - 175 cc/min
  • Span gas - methane is typical

Display Width

N/A 3.4 in8.64 cm

Display Length

N/A 4.5 in11.43 cm

Display Type

N/A Graphical Liquid Crystal Display (LCD)

Outputs

N/A
  • LAN/Ethernet, RS-232
  • Analog, 1 programmable 0-20mA/4-20mA isolated output
  • Relays, 5 programmable From A relays rated to 3A at 230VAC

Operating Temperature

N/A 32 to 104 ºF0 to 40 ºC

Non Condensing Operating Humidity

N/A 0 to 95 %

Prevenitive Maintenance Kit

N/A Includes all parts potentially needed on-hand

Gas Generators

N/A
  • Hydrogen Generator
  • Zero Air generators

Software

N/A 9000 Keeper provided for method switching via PC

Service and Support Options

N/A Start-up and training, 1 day onsite

Height

N/A 5.25 in13.02 cm

Width

N/A 19 in48.26 cm

Depth

N/A 16.25 in41.28 cm

Weight

N/A <20 lb9.07 kg

Alternating Current (AC) Voltage

N/A 100 to 240 V

Frequency

N/A 50 Hz60 Hz

Current

N/A 1 A

Configuration

Configuration

N/A Benchtop or 19" (48.3 cm) Rack-Mount, 3U

Applications

Applications

N/A
  • Compliance monitoring for EPA Method 25A
  • Continuous emission stack monitoring
  • Scrubber and oxidizer efficiency
  • Carbon bed breakthrough detection
  • Vehicle emissions

Features

Features

N/A
  • Automatic FID ignition
  • Automatic calibration at user-defined intervals
  • Internal multi-point sampling option
  • Electronic back-pressure regulator with sample bypass system
  • Flow Guard Control of fuel, air, and sample
  • Graphic LCD display with easy to use menu
  • Benchtop or rack-mountable
  • Single or multi-point sampling
  • Customizable Ethernet or serial output
  • Programmable analog output ranges
  • Programmable relays for iagnostics, concentration, alarms, and events

Additional Information

Additional Information

N/A All instrument parameters are reported clearly and continually refreshed on a large, graphical LCD display. Using analog, digital, and logic output communication capabilities, analytical information from the analyzer can be acquired using an external PC and a simple communications program such as Windows® HyperTerminal or the analyzer can output binary or ASCII formats directly to a data acquisition system or PLC. Every Baseline® 9000 analyzer includes MOCON’s free PC utility 9000 Keeper used for storing and uploading multiple methods, as well as sending configuration settings, directly to the analyzer.
  • 1 Fuel blend options available